Table 1 Effect of dynamic range, gain and probe distance on lumen diameter measurements

From: Ultrasound settings significantly alter arterial lumen and wall thickness measurements

  

Phantom A

 

Phantom B

 

Phantom C

 
  

Lumen (mm)

p

Lumen (mm)

p

Lumen (mm)

p

Dynamic range (dB)

40

4.77 ± 0.09

<0.001

5.71 ± 0.09

<0.001

4.75 ± 0.10

<0.001

 

45

4.75 ± 0.09

 

5.69 ± 0.09

 

4.73 ± 0.10

 
 

50

4.73 ± 0.08

 

5.68 ± 0.08

 

4.71 ± 0.09

 
 

55

4.71 ± 0.08

 

5.66 ± 0.07

 

4.68 ± 0.09

 
 

60

4.70 ± 0.07

 

5.64 ± 0.06

 

4.67 ± 0.08

 

Gain (dB)

-10

4.79 ± 0.07

<0.001

5.75 ± 0.06

<0.001

4.77 ± 0.07

<0.001

 

-5

4.77 ± 0.07

 

5.71 ± 0.06

 

4.75 ± 0.08

 
 

0

4.74 ± 0.08

 

5.68 ± 0.06

 

4.71 ± 0.08

 
 

+5

4.70 ± 0.08

 

5.64 ± 0.06

 

4.67 ± 0.09

 
 

+10

4.67 ± 0.08

 

5.60 ± 0.06

 

4.64 ± 0.09

 

Probe distance (mm)

10

4.80 ± 0.07

<0.001

5.74 ± 0.06

<0.001

4.75 ± 0.08

<0.001

 

15

4.75 ± 0.08

 

5.71 ± 0.07

 

4.75 ± 0.07

 
 

20

4.70 ± 0.09

 

5.64 ± 0.07

 

4.67 ± 0.10

 
 

25

4.68 ± 0.08

 

5.62 ± 0.06

 

4.67 ± 0.10

 
  1. Data are mean ± standard deviation
  2. n = 80 per level for dynamic range (5 gain levels × 4 distance levels × 4 angles)
  3. n = 80 per level for gain (5 dynamic range levels × 4 distance levels × 4 angles)
  4. n = 100 per level for probe distance (5 gain levels × 5 dynamic range levels × 4 angles)
  5. P-values are for trend from a univariate linear regression model
  6. Phantom A and C have an estimated lumen diameter of 4.98 ± 0.01 mm
  7. Phantom B has an estimated lumen diameter of 5.98 ± 0.01 mm.