Skip to main content

Table 6 Comparison of 3-dimensional left ventricular strains using two different software (4D LV analysis: version 3.1, EchoPAC: version 12.0)

From: The impact of preload on 3-dimensional deformation parameters: principal strain, twist and torsion

 

4D LV analysis

EchoPAC

P value

ICC

95% CI

Pre-HD (n = 58)

 Global longitudinal strain (%)

−24.57 ± 2.51

−17.10 ± 2.34

<0.001

−0.089

−0.840-0.355

 Global circumferential strain (%)

−33.35 ± 3.50

−19.95 ± 2.51

<0.001

0.408

0.000–0.649

 Global radial strain (%)

47.67 ± 4.27

53.45 ± 9.64

<0.001

−0.055

−0.783-0.376

Post-HD (n = 58)

 Global longitudinal strain (%)

−21.42 ± 2.15

−15.58 ± 3.21

<0.001

0.116

−0.494-0.477

 Global circumferential strain (%)

−30.90 ± 3.22

−17.42 ± 3.71

<0.001

0.570

0.273–0.745

 Global radial strain (%)

42.90 ± 3.61

45.79 ± 11.31

0.045

0.304

−0.176-0.588

  1. ICC intra-class correlation coefficient, CI confidence interval, HD hemodialysis