Skip to main content

Table 5 Multivariate models for prediction of sub-optimal MitraClip response

From: Transcatheter MitraClip repair alters mitral annular geometry – device induced annular remodeling on three-dimensional echocardiography predicts therapeutic response

5A.Multivariate Regression
Model chi square = 17.22, p < 0.001
VariableOdds Ratio (95% Confidence Interval)p
Mitral Annular Area (cm2/m2)1.93 (1.19–3.13)0.007
LV End Diastolic Volume (ml/m2)a1.29 (1.02–1.63)0.03
5B.Multivariate Regression
Model chi square = 10.87, p = 0.004
VariableOdds Ratio (95% Confidence Interval)p
Mitral Annular Diameter (cm/m2)5.36 (0.95–30.19)0.06
LV End Diastolic Volume (ml/m2)a1.23 (1.02–1.49)0.03
  1. aper 10 ml increment
  2. Data in bold are significant