Table 2 Comparison of maximum diameter, vessel area, and vessel volume

From: Feasibility and accuracy of a novel automated three-dimensional ultrasonographic analysis system for abdominal aortic aneurysm: comparison with two-dimensional ultrasonography and computed tomography

Variable

Overall (n = 77)

Control subjects (n = 18)

Patients with AAA (n = 59)

CT

 Dmax, mm

35.4 ± 12.0

18.9 ± 2.5

40.1 ± 9.0

 Vessel area, cm2

10.9 ± 7.1

2.4 ± 0.6

13.3 ± 6.2

 Vessel volume, cm3

44.4 ± 36.1

2.1 ± 0.5

57.3 ± 31.4

2-D US

 Dmax, mm

32.3 ± 12.1

15.0 ± 2.6

37.6 ± 8.3

 Vessel area, cm2

10.3 ± 7.1

1.9 ± 0.4

12.8 ± 6.1

3-D US

 Dmax, mm

35.1 ± 12.0

19.0 ± 2.7

39.9 ± 9.2

 Vessel area, cm2

10.6 ± 7.0

2.1 ± 0.4

13.1 ± 6.7

 Vessel volume, cm3

42.4 ± 33.9

1.9 ± 0.4

54.0 ± 29.0

  1. AAA abdominal aortic aneurysm, CT computed tomography, Dmax maximum diameter of the abdominal aorta, 2-D US two-dimensional ultrasonography, 3-D US three-dimensional ultrasonography