Determinants of clinical improvement after surgical replacement or transcatheter aortic valve implantation for isolated aortic stenosis

Abstract

Background

Transcatheter aortic valve implantation (TAVI) is an alternative to surgical aortic valve replacement (SAVR) in patients with aortic stenosis (AS) and high surgical risk. Hemodynamic performance after TAVI is superior, but the impact of reverse remodeling on clinical improvement is controversial. We aim to address the differences in hemodynamic changes between SAVR and TAVI, and its correlation with LV remodeling and clinical improvement at 6 months follow-up.

Methods

Forty-two patients treated by TAVI were compared with 45 SAVR patients with a stented bioprosthesis. Clinical, 2D and 3D echocardiographic data were prospectively obtained before and six months after intervention.

Results

Patients had similar distribution for sex, body surface area and AS severity. TAVI patients were older, more symptomatic and had more comorbidities. They also had higher LV filling pressures, larger 3D indexed left atrium volume, but similar 3D indexed LV mass. At 6 months, TAVI patients had greater clinical improvement and higher effective orifice area index (EAOI), but only SAVR patients already had a significant decrease in 3D indexed LV mass and diastolic volume. In univariate analysis older age, NYHA class ≥ III, increase in EAOI and TAVI were related with functional class improvement. After multivariate analysis only NYHA class ≥ III (OR 8.81, CI:2.13-36.52; p = 0.003) and an increase in EAOI ≥ 105% (OR 3.87, CI:1.02-14.70; p = 0.04) were predictors of clinical improvement.

Conclusions

At 6 months, functional class improvement was greater after TAVI. Higher initial NYHA class and an increase in EAOI ≥ 105% were independently associated with functional enhancement. It is debatable if left ventricular remodeling is determinant for functional class improvement.

Peer Review reports

Background

Aortic stenosis (AS) is the most prevalent of all valvular diseases in developed countries and its increase has a direct relation with population aging [1]. In elderly patients cardiac surgery can be challenging by the increased number of comorbidities, making transcatheter aortic valve implantation (TAVI) an atractive alternative treatment modality [2].

Surgical aortic valve replacement (SAVR) nearly normalizes long-term survival and improves quality of life in AS patients [3, 4] but late outcomes depend mainly on the stage of heart disease before surgery, prosthetic related complications, and comorbidities. Although there is a significant reduction of wall stress and left ventricular (LV) pressure after SAVR, nearly half of patients have residual LV hypertrophy (LVH) late after surgery [5, 6]. This persistent increase in LV mass is an independent predictor of cardiac-related morbidity [6] and mortality [7] making reverse remodeling an important outcome after surgery.

Similarly, TAVI has shown good mid-term results, not inferior to SAVR in high-risk patients [2] and superior to medical therapy [8]. TAVI patients have higher effective orifice areas (EOA) and lower transprosthetic gradients but, in spite of remarkable clinical improvement, LV mass regression and reverse remodeling are less consistent in comparison with SAVR [911].

In this study we aim to address the differences in hemodynamic changes between SAVR and TAVI patients, and its correlation with LV remodeling and clinical improvement at 6 months follow-up.

Methods

Patient selection and follow-up

The present study is a comparison of two prospective cohorts of patients with symptomatic degenerative severe AS (defined as aortic valve area ≤1 cm2) and LV ejection fraction (EF) ≥ 40%, who underwent SAVR with a stented bioprosthesis or TAVI. This was a collaborative work from two distinct institutions since one of them didn’t have a TAVI program at the time of patient inclusion.

Patients with aortic regurgitation > II/IV, moderate to severe mitral or tricuspid regurgitation, significant coronary artery disease (lesions >50% on coronary angiography) or previous cardiac surgery were excluded. All patients had a clinical and echocardiographic evaluation before and at 6 months after the procedure, if alive.

This investigation conforms to the Declaration of Helsinki, had institutional ethical review board approval and each study participant signed an informed consent before enrolment.

TAVI group

Forty-two consecutive patients with severe aortic stenosis and preserved LV systolic function, submitted to successful TAVI in one tertiary center, Hospital Clinico San Carlos, from April 2009 to April 2010, were included. These patients were obtained from a series of 97 consecutive patients who underwent TAVI, after excluding those with significant LV systolic dysfunction, concomitant moderate to severe mitral valve disease or aortic regurgitation, and those with significant coronary artery disease.

Patients were referred for TAVI due to an excessive risk for SAVR, which was estimated using the logistic EuroSCORE and/or clinical judgment.

Vascular access was obtained either by percutaneous approach through the common femoral artery (30 patients) or transapical approach (12 patients).

The procedure was performed under fluoroscopy and transesophageal echocardiography (TEE) guidance using the techniques described in detail in previous reports [12]. Among all, 31 (73.8%) patients were implanted with an Edwards SAPIEN (Edwards Lifesciences, Irvine, CA, USA). The CoreValve (Medtronic CoreValve Percutaneous System, Medtronic CV) was implanted in 11 (26.2%) patients exclusively by retrograde transfemoral approach. Two valve sizes were available, 23- and 26-mm expanded diameter for Edwards SAPIEN valve and 26- and 29-mm for CoreValve. The prosthesis size was decided according to annulus diameter, measured by TEE. The deployment was performed under the agreement of the interventionist and the echocardiographer. Device success was defined as stable device placement and function as assessed by angiography and echocardiography. All patients with developing new grade III atrioventricular block were implanted with a permanent pacemaker within 3 days after valve implantation.

SAVR group

Between January 2009 and December 2009, among 141 consecutive patients with isolated symptomatic AS referred for SAVR at the Cardiothoracic Surgery Department of Centro Hospitalar São João, Porto, Portugal we included 45 patients with 3D echocardiographic evaluation. All surgeries were performed using standard procedure for aortic valve replacement. Patients were placed on cardiopulmonary bypass and cardiac arrest was induced and maintained with cold blood cardioplegia. The prosthetic substitutes used in this study were Carpentier-Edwards Perimount pericardial valve (Edwards Lifesciences, Irvine, CA) and the St Jude Medical Epic porcine valve (St Jude Medical, Inc, St Paul, Minn). Valve sizes were 19 mm (n = 4;8.9%), 21 mm (n = 25;55.6%), 23 mm (n = 11;24.4%) and 25 mm (n = 5;11.1%).

Echocardiographic studies

Echocardiographic examination was performed by a trained cardiologist and recorded on digital support. All recordings were examined by an experienced echocardiographer in an accredited independent echocardiography laboratory (Hospital Clínico San Carlos in Madrid, Spain), blinded to patient details. Studies were performed using Phillips iE-33 equipment with a S5-1 transducer with M-mode, two dimensional, pulsed, continuous, color-flow and tissue Doppler capabilities, and an X3-1 transducer for 3D imaging. Imaging analysis was performed with Xcelera and QLab software. All measurements were performed in accordance with the recommendations of the American Society of Echocardiography [13]. Peak transvalvular gradient was estimated using the simplified Bernoulli equation. Aortic valve area (or effective orifice area, EOA) was estimated using quantitative Doppler by the continuity equation. The EOA values were then indexed to body surface area (EAOI). Patient prosthesis mismatch was considered present if EAOI ≤0.85 cm2/m2 and severe when EAOI ≤ 0.65 cm2/m2.

The presence, degree, and type (paravalvular vs transvalvular) of aortic regurgitation (AR) were classified as trivial, mild, moderate, or severe according to The Valve Academic Research Consortium (VARC) II [14].

Mitral inflow was assessed in the apical 4-chamber view using pulsed wave Doppler with the sample volume placed at the tips of mitral leaflets during diastole. From the mitral inflow profile, the peak flow velocity of early filling (E wave), peak flow velocity of atrial contraction (A wave), the E/A ratio, and early filling deceleration time (DT) were measured. Doppler tissue imaging of the mitral annulus was obtained from the apical 4-chamber using a sample volume placed in the septal mitral valve annulus. The septal e’ velocity value was determined, and the E/e’ ratio was derived.

Left atrium (LA) volume, LV systolic and diastolic volumes, and resulting ejection fraction were calculated with direct 3D volumetric analysis.

All indexed values were obtained by dividing by body surface area according to the formula of Mosteler.

Statistical analysis

Categorical variables were expressed as percentages and continuous variables as mean ± standard deviation or median and interquartile range, according to their distribution. Continuous variables were compared between groups using an unpaired t-test (for normally distributed variables) or the Mann–Whitney U-test (for non-normally distributed variables). For comparison between baseline and follow-up a paired Student's t-test was applied (normally distributed variables) and a Wilcoxon test (for non-normal distributed variables). Chi-square test was used to compare proportions. Following univariate analysis, a stepwise binary logistic multivariate regression model (Wald backward stepwise method, p = 0.05 for covariate inclusion and 0.1 for exclusion) was performed, including potential confounders for NYHA improvement regression analysis 6 months after AVR. NYHA improvement was analyzed as worse or equal vs better. Of note, patients in baseline NYHA class I but symptomatic (angina or syncope) were considered improved if these additional symptoms disappeared.

All reported probability values are two-tailed, and p < 0.05 was considered statistically significant. Analyses were performed with the IBM®SPSS® Statistics software package (version 21.0) (SPSS Inc, Chicago, IL, USA).

Results

Patient’s baseline clinical characteristics are described in Table 1. TAVI patients were older, had more comorbidities and higher logistic Euroscore. Patients who underwent TAVI had worse functional class at baseline, although they had similar baseline severity of aortic stenosis (Table 2). Before intervention, when comparing to SAVR patients, TAVI patients had similar 3D LV mass index but, after normalizing LV mass to LV end-diastolic volume (LVMI/LVDVI), they had more concentric geometry (3.0 (P25-75:2.1-4.4) vs 2.4 (P25-75:1.8-3.0) g/ml; p = 0.044) due to smaller LV end-diastolic index volumes. TAVI patients also had worse diastolic dysfunction, with higher LV filling pressures and larger indexed LA volume (Table 3).

Table 1 Baseline clinical characteristics of SAVR vs TAVI patients
Table 2 Baseline and 6 months 2D echocardiographic data on SAVR vs TAVI patients
Table 3 Baseline and 6 months 3D echocardiographic data on SAVR vs TAVI patients

At 6 months follow-up, 5 TAVI patients had died, 3 during hospitalization for TAVI, and 2 after hospital discharge from non-cardiovascular causes. There were no deaths in the SAVR group.

Changes in LV remodeling and functional class after aortic valve replacement

At 6 months (Table 2), TAVI patients had a higher effective orifice area index (EAOI) and lower transprosthetic maximal velocity and mean gradient, as well as a greater absolute increase in EAOI. Patient-prosthesis mismatch (PPM) was more frequent in the SAVR patients and there were no severe PPM cases in the TAVI group. There was a significant increase in ejection fraction (EF) in both groups and, when considering LV remodeling (Table 3), although there was a decrease in LV mass index (LVMI) and LV diastolic volume index (LVDVI) in both groups, only in SAVR patients this decrease was significant when compared with baseline values (Figure 1).

Figure 1
figure 1

Six months changes in parameters of remodeling in TAVI and SAVR groups. A- change in left ventricular mass index (LVMI); B- change in ratio LVMI/left ventricular end-diastolic volume index (LVDVI); C- change in ejection fraction (EF); D- change in left atrial volume index (LAVI); E- change in LVDVI; F- change in left ventricular end-systolic volume index (LVSVI).

The presence of patient-prosthesis mismatch had no correlation with changes in LVMI, LVDVI, LV end-systolic volume index (LVSVI), or LA volume index (Additional file 1: Table S1). Moreover, EAOI increase was not related to final LVMI (r = 0.082, p = 0.512), LVDVI (r = 0.015, p = 0.925), LVSVI (r = 0.154, p = 0.331), or LAVI (r = 0.187, p = 0.143).

After intervention, NYHA class was better in 30 (71.4%) TAVI patients compared with 22 (48.8%) SAVR patients (p = 0.001, Figure 2). Patients exhibiting a better NYHA class were more likely to have no PPM (81.6% vs 60.0%, p = 0.049) and a greater relative increase in EAOI (163.2% (P25-75: 118.9-234.8) vs 103.0% (P25-75: 52.1-170.2); p = 0.030). In univariate analysis, older age, baseline NYHA class ≥ III, a higher increase in EAOI and TAVI procedure were related with functional class improvement (Table 4). Moreover, we found no correlation between functional class improvement and parameters of LV reverse remodeling like the decrease in LV volumes or mass, or improvement in diastolic function suggested by the decrease in left atrial volume and E/e’ (Table 4). After a stepwise logistic multivariate regression analysis, the only independent predictors of NYHA class improvement were initial NYHA class ≥ III (OR 8.81, CI: 2.13-36.52; p = 0.003) and relative increase in EAOI ≥ 105% (OR 3.87, CI: 1.02-14.70; p = 0.04).

Figure 2
figure 2

Six months changes in NYHA class in TAVI and SAVR groups.

Table 4 Univariate and multivariate analysis of clinical and echocardiographic determinants of NYHA class improvement

Paravalvular regurgitation was more frequent after TAVI (59% vs 6.7%, p < 0.001), mostly of mild degree. Only 5 TAVI patients (5.8%) had moderate aortic regurgitation at six months. Paravalvular regurgitation had no correlation with the variation of indexed LVM, LVDV or LVSV (Additional file 2: Table S2). NYHA class improvement was similar in patients with and without paravalvular regurgitation (83.3% vs 66.7%, p = 0.241).

Discussion

In this study we found that, at 6 months, TAVI patients had a better hemodynamic result and greater clinical improvement than those submitted to SAVR, but LV reverse remodeling was of a less significant degree than in SAVR patients.

Reverse remodeling after aortic valve replacement

At 6 months follow-up, TAVI patients had a more favorable hemodynamic performance but the decrease in LVMI and LV volumes, although showing the same trend as SAVR patients, was less extensive.

A greater decrease in transvalvular gradients and increase in effective orifice area was seen after TAVI. It would be expected that patients undergoing this procedure had faster remodeling if load was the most important determinant of mass regression, but in our study EAOI increase was not associated with changes in LV mass and volumes after AVR. We can speculate that the baseline differences we have found, with older age and more comorbidities in TAVI patients, could have contributed to this result. Moreover, TAVI patients were “sicker”, with worse diastolic dysfunction and worse functional class, despite similar severity of AS and EF, possibly due to longer time of LV overload exposure. This could explain a restricted ability of the myocardium to recover from pre-intervention changes.

Various groups have focused on the hemodynamic and structural effects of TAVI [10, 1518] with consistent results in afterload reduction and symptomatic improvement, but with conflicting results on reverse remodeling. Previous reports comparing the impact of TAVI and SAVR on LV remodeling addressed heterogeneous populations, including patients with coronary artery disease, different levels of EF, and several types of aortic prosthesis, including mechanical, stented and stentless bioprosthesis [9, 11, 19].

Clavel et al. [20] compared hemodynamic performances of TAVI or SAVR with stentless and stented bioprosthesis. At 6 to 12 months the increase in EAOI and reduction in transvalvular gradients in TAVI patients was similar to that obtained with stentless bioprosthesis and there was a clear advantage of TAVI in preventing PPM in patients with small annulus (≤20 mm). However, data on clinical improvement or reverse remodeling is absent.

Fairbairn et al. [19], using cardiac magnetic resonance imaging, showed a decrease in LVMI and indexed LV systolic volume at 6 months in both groups, but only SAVR patients had a decrease in indexed LV diastolic volume. The authors considered that the smaller reduction in LV end-diastolic volume post-TAVI could be related to a greater burden of coronary artery disease in these patients. In our study we found that, at 6 months, only SAVR patients had a significant decrease in LV mass and LV end-diastolic volume. Since we excluded patients with coronary artery disease, the absence of significant remodeling post-TAVI at this time point could indicate the existence of irreversible disease. Constantino et al. [18] compared reverse remodeling 2 months after TAVI and SAVR, and concluded that there was a more significant reduction in LVMI and relative wall thickness (RWT) in TAVI patients. These results are conflicting with ours, but were obtained in an earlier time point and the lack of adjustment for differences in baseline LVMI could have influenced results. Moreover, the authors found that these structural changes were paralleled by reduction in estimated filling pressures after TAVI. We also found a reduction in E/e´ after TAVI, but this occurred in absence of favorable remodeling and had no correlation with clinical improvement. Its association with prognosis is yet to be seen.

In the Cohort A of the PARTNER trial, LV diastolic volume was higher in TAVI patients in the first year of follow-up, but these differences were no longer present at two years. LV mass regression was faster in the SAVR group, although there were no significant differences after 6 months [11]. These results are similar to those reported in our study, suggesting that reverse remodeling can also occur in TAVI patients, but the process is slower than after SAVR, even after matching for age and major comorbidities.

Aortic regurgitation after TAVI, mostly paravalvular, is a common event [2, 21] and has come to our attention because of its impact on mortality. Post-procedural moderate to severe AR increased in-hospital mortality in comparison with no or only mild AR [21] and, in the randomized PARTNER trial, there was a positive correlation between AR severity and long-term mortality [22]. The pathophysiology underlying this increase is mortality is unclear. It has been speculated that significant paravalvular aortic regurgitation can overload the LV and impair reverse remodeling [23], therefore worsening prognosis. In our study, the presence of paravalvular aortic regurgitation had no correlation with the variation in LV volumes or mass. Once only 5 patients had moderate regurgitation, the lack of association with ventricular remodeling could be due to the small sample size. Longer follow-up and larger numbers are needed to take any definite conclusion on its impact in remodeling.

Predictors of clinical improvement

In our sample, the increase in EAOI to more than the double was a strong predictor for clinical improvement, independent of changes in parameters of reverse remodeling. Conversely, the presence of patient-prosthesis mismatch (PPM) was correlated with impaired improvement in NYHA class.

Several studies reported that PPM is an independent predictor of cardiac events after AVR [24, 25] while others failed to demonstrate a significant impact on outcomes [26, 27].

Some authors found that persistent PPM results in less regression [28, 29] but even patients with PPM or small prosthesis can have significant reduction in LVM [30, 31]. The extent of regression is largely dependent on the extent of EOA increase after AVR [32]. Given the curvilinear relation of indexed EOA and transprosthetic gradients the degree of regression seems to be dependent on the original and final positions of an individual patient on the indexed EOA-gradient curve [33]. Although we found no correlation between PPM and impaired LVMI regression, we did find that the increase in valve area to more than twice the initial value was crucial for clinical improvement. This can be particularly important in elderly patients whose main concern is the achievement of a better quality of life.

Finally, as expected, patients who were in worse NYHA class before intervention more frequently experienced a clinical improvement. Using NYHA class to evaluate clinical improvement, although extensively used, has limitations and it is easier to demonstrate an improvement when a patient is class III/IV than NYHA class II/I. This fact can also help to explain the better improvement in the clinical status of patients undergoing TAVI, as they were in worse NYHA class than SAVR patients, before the intervention.

Limitations

Since this was an observational study, we were not able to match patients for age, comorbidities or prosthesis size. These factors were considered in the multivariate model used for prediction of clinical improvement, but the authors recognize that, although logistic multivariate analysis is commonly used, it can’t correct for all possible confounders.

In addition, we selected our population by excluding patients with concomitant coronary artery disease and significant associated valvular disease to reduce introduction of further bias. The limited number of patients included in this analysis has limited statistical power to detect small differences between groups.

The evaluation of functional improvement using NYHA class is subjective and a more objective method, like six-minute walk test, could have allowed a quantitative assessment. Nevertheless, in clinical practice, NYHA class is the most widely use classification of function status and has been proven useful over the years.

Conclusions

At six months after aortic valve intervention, better hemodynamic result was seen after TAVI, but LV reverse remodeling was of a less significant degree than after SAVR. Older age, comorbidities and the existence of a more extensive myocardial disease, as suggested by worse diastolic dysfunction and worse functional class in TAVI patients, despite similar severity of AS and EF, could explain a restricted ability of the myocardium to recover even after load relief. Moreover, six months may be too early to draw definitive conclusions, namely regarding the consequences of paravalvular aortic regurgitation.

Mid-term clinical improvement was strongly related to the increase in EAOI and had no association with LV remodeling parameters. Thus, doubling the initial aortic valve area seems to be a key point to achieve clinical improvement after valve replacement, a particularly important endpoint in the elderly.

This study raises some important new questions but longer follow-up and large-scale randomized trials are needed to confirm these results.

Acknowledgements

This work was supported by the Cardiovascular R&D Center, through grants from the Portuguese Foundation for Science and Technology (PEst-C/SAU/UI0051/2011, EXCL/BIM-MEC/0055/2012 and HMSP-ICS/0007/2012; partially funded by FEDER through COMPETE) and from the European Commission (FP7-Health-2010; MEDIA-261409).

Abbreviations

SAVR:

Surgical aortic valve replacement

TAVI:

transcatheter aortic valve implantation

BSA:

body surface area

COPD:

chronic obstructive pulmonary disease

PAD:

peripheral artery disease

DM:

diabetes mellitus

HT:

arterial hypertension

NYHA:

New York Heart Association

AV:

aortic valve

AVA:

aortic valve area

EAOI:

effective orifice area index

PPM:

patient-prosthesis mismatch

IVRT:

Isovolumetric relaxation time

LVDVI:

left ventricular end-diastolic volume index

LVSVI:

left ventricular end-systolic volume index

LVMI:

left ventricular mass index

LAVI:

left atrial volume index.

References

  1. Supino PG, Borer JS, Preibisz J, Bornstein A: The epidemiology of valvular heart disease: A growing public health problem. Heart failure clinics. 2006, 2: 379-393. 10.1016/j.hfc.2006.09.010.

    Article  PubMed  Google Scholar 

  2. Smith CR, Leon MB, Mack MJ, Miller DC, Moses JW, Svensson LG, Tuzcu EM, Webb JG, Fontana GP, Makkar RR, Williams M, Dewey T, Kapadia S, Babaliaros V, Thourani VH, Corso P, Pichard AD, Bavaria JE, Herrmann HC, Akin JJ, Anderson WN, Wang D, Pocock SJ, Investigators PT: Transcatheter versus surgical aortic-valve replacement in high-risk patients. The New England journal of medicine. 2011, 364: 2187-2198. 10.1056/NEJMoa1103510.

    Article  CAS  PubMed  Google Scholar 

  3. Schwarz F, Baumann P, Manthey J, Hoffmann M, Schuler G, Mehmel HC, Schmitz W, Kubler W: The effect of aortic valve replacement on survival. Circulation. 1982, 66: 1105-1110. 10.1161/01.CIR.66.5.1105.

    Article  CAS  PubMed  Google Scholar 

  4. Kvidal P, Bergstrom R, Horte LG, Stahle E: Observed and relative survival after aortic valve replacement. Journal of the American College of Cardiology. 2000, 35: 747-756. 10.1016/S0735-1097(99)00584-7.

    Article  CAS  PubMed  Google Scholar 

  5. Lund O, Kristensen LH, Baandrup U, Hansen OK, Nielsen TT, Emmertsen K, Jensen FT, Flo C, Rasmussen BS, Pilegaard HK: Myocardial structure as a determinant of pre- and postoperative ventricular function and long-term prognosis after valve replacement for aortic stenosis. European heart journal. 1998, 19: 1099-1108. 10.1053/euhj.1998.0872.

    Article  CAS  PubMed  Google Scholar 

  6. Zybach-Benz RE, Aeschbacher BC, Schwerzmann M: Impact of left ventricular hypertrophy late after aortic valve replacement for aortic stenosis on cardiovascular morbidity and mortality. International journal of cardiology. 2006, 109: 41-47. 10.1016/j.ijcard.2005.05.050.

    Article  PubMed  Google Scholar 

  7. Lund O: Preoperative risk evaluation and stratification of long-term survival after valve replacement for aortic stenosis. Reasons for earlier operative intervention. Circulation. 1990, 82: 124-139.

    CAS  PubMed  Google Scholar 

  8. Leon MB, Smith CR, Mack M, Miller DC, Moses JW, Svensson LG, Tuzcu EM, Webb JG, Fontana GP, Makkar RR, Brown DL, Block PC, Guyton RA, Pichard AD, Bavaria JE, Herrmann HC, Douglas PS, Petersen JL, Akin JJ, Anderson WN, Wang D, Pocock S, Investigators PT: Transcatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery. The New England journal of medicine. 2010, 363: 1597-1607. 10.1056/NEJMoa1008232.

    Article  CAS  PubMed  Google Scholar 

  9. Giannini C, Petronio AS, Nardi C, De Carlo M, Guarracino F, Delle Donne MG, Talini E, Minzioni G, Bortolotti U, Cucco C, Marzilli M, Di Bello V: Left ventricular reverse remodeling in percutaneous and surgical aortic bioprostheses: An echocardiographic study. Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography. 2011, 24: 28-36. 10.1016/j.echo.2010.08.027.

    Article  Google Scholar 

  10. Sherif MA, Abdel-Wahab M, Awad O, Geist V, El-Shahed G, Semmler R, Tawfik M, Khattab AA, Richardt D, Richardt G, Tolg R: Early hemodynamic and neurohormonal response after transcatheter aortic valve implantation. American heart journal. 2010, 160: 862-869. 10.1016/j.ahj.2010.07.017.

    Article  PubMed  Google Scholar 

  11. Hahn RT, Pibarot P, Stewart WJ, Weissman NJ, Gopalakrishnan D, Keane MG, Anwaruddin S, Wang Z, Bilsker M, Lindman BR, Herrmann HC, Kodali SK, Makkar R, Thourani VH, Svensson LG, Akin JJ, Anderson WN, Leon MB, Douglas PS: Comparison of transcatheter and surgical aortic valve replacement in severe aortic stenosis: A longitudinal study of echocardiography parameters in cohort a of the partner trial (placement of aortic transcatheter valves). Journal of the American College of Cardiology. 2013, 61: 2514-2521. 10.1016/j.jacc.2013.02.087.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Zamorano JL, Badano LP, Bruce C, Chan KL, Goncalves A, Hahn RT, Keane MG, La Canna G, Monaghan MJ, Nihoyannopoulos P, Silvestry FE, Vanoverschelde JL, Gillam LD: Eae/ase recommendations for the use of echocardiography in new transcatheter interventions for valvular heart disease. European heart journal. 2011, 32: 2189-2214. 10.1093/eurheartj/ehr259.

    Article  PubMed  Google Scholar 

  13. Lang RM, Bierig M, Devereux RB, Flachskampf FA, Foster E, Pellikka PA, Picard MH, Roman MJ, Seward J, Shanewise JS, Solomon SD, Spencer KT, Sutton MS, Stewart WJ: Recommendations for chamber quantification: A report from the american society of echocardiography's guidelines and standards committee and the chamber quantification writing group, developed in conjunction with the european association of echocardiography, a branch of the european society of cardiology. Journal of the American Society of Echocardiography: official publication of the American Society of Echocardiography. 2005, 18: 1440-1463. 10.1016/j.echo.2005.10.005.

    Article  Google Scholar 

  14. Holmes DR, Mack MJ, Kaul S, Agnihotri A, Alexander KP, Bailey SR, Calhoon JH, Carabello BA, Desai MY, Edwards FH, Francis GS, Gardner TJ, Kappetein AP, Linderbaum JA, Mukherjee C, Mukherjee D, Otto CM, Ruiz CE, Sacco RL, Smith D, Thomas JD: 2012 accf/aats/scai/sts expert consensus document on transcatheter aortic valve replacement. Journal of the American College of Cardiology. 2012, 59: 1200-1254. 10.1016/j.jacc.2012.01.001.

    Article  PubMed  Google Scholar 

  15. La Manna A, Sanfilippo A, Capodanno D, Salemi A, Cadoni A, Cascone I, Polizzi G, Figuera M, Pittala R, Privitera C, Tamburino C: Left ventricular reverse remodeling after transcatheter aortic valve implantation: A cardiovascular magnetic resonance study. Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance. 2013, 15: 39-10.1186/1532-429X-15-39.

    Article  Google Scholar 

  16. Tzikas A, Geleijnse ML, Van Mieghem NM, Schultz CJ, Nuis RJ, van Dalen BM, Sarno G, van Domburg RT, Serruys PW, de Jaegere PP: Left ventricular mass regression one year after transcatheter aortic valve implantation. The Annals of thoracic surgery. 2011, 91: 685-691. 10.1016/j.athoracsur.2010.09.037.

    Article  PubMed  Google Scholar 

  17. Gotzmann M, Lindstaedt M, Bojara W, Mugge A, Germing A: Hemodynamic results and changes in myocardial function after transcatheter aortic valve implantation. American heart journal. 2010, 159: 926-932. 10.1016/j.ahj.2010.02.030.

    Article  PubMed  Google Scholar 

  18. Costantino MF, Galderisi M, Dores E, Innelli P, Tarsia G, Di Natale M, Santoro C, De Stefano F, Esposito R, de Simone G: Parallel improvement of left ventricular geometry and filling pressure after transcatheter aortic valve implantation in high risk aortic stenosis: Comparison with major prosthetic surgery by standard echo doppler evaluation. Cardiovascular ultrasound. 2013, 11: 18-10.1186/1476-7120-11-18.

    Article  PubMed  PubMed Central  Google Scholar 

  19. Fairbairn TA, Steadman CD, Mather AN, Motwani M, Blackman DJ, Plein S, McCann GP, Greenwood JP: Assessment of valve haemodynamics, reverse ventricular remodelling and myocardial fibrosis following transcatheter aortic valve implantation compared to surgical aortic valve replacement: A cardiovascular magnetic resonance study. Heart. 2013, 99: 1185-1191. 10.1136/heartjnl-2013-303927.

    Article  PubMed  PubMed Central  Google Scholar 

  20. Clavel MA, Webb JG, Pibarot P, Altwegg L, Dumont E, Thompson C, De Larochelliere R, Doyle D, Masson JB, Bergeron S, Bertrand OF, Rodes-Cabau J: Comparison of the hemodynamic performance of percutaneous and surgical bioprostheses for the treatment of severe aortic stenosis. Journal of the American College of Cardiology. 2009, 53: 1883-1891. 10.1016/j.jacc.2009.01.060.

    Article  PubMed  Google Scholar 

  21. Abdel-Wahab M, Zahn R, Horack M, Gerckens U, Schuler G, Sievert H, Eggebrecht H, Senges J, Richardt G: German transcatheter aortic valve interventions registry i. Aortic regurgitation after transcatheter aortic valve implantation: Incidence and early outcome. Results from the german transcatheter aortic valve interventions registry. Heart. 2011, 97: 899-906.

    PubMed  Google Scholar 

  22. Kodali SK, Williams MR, Smith CR, Svensson LG, Webb JG, Makkar RR, Fontana GP, Dewey TM, Thourani VH, Pichard AD, Fischbein M, Szeto WY, Lim S, Greason KL, Teirstein PS, Malaisrie SC, Douglas PS, Hahn RT, Whisenant B, Zajarias A, Wang D, Akin JJ, Anderson WN, Leon MB, Investigators PT: Two-year outcomes after transcatheter or surgical aortic-valve replacement. The New England journal of medicine. 2012, 366: 1686-1695. 10.1056/NEJMoa1200384.

    Article  CAS  PubMed  Google Scholar 

  23. Merten C, Beurich HW, Zachow D, Mostafa AE, Geist V, Toelg R, Richardt G, Abdel-Wahab M: Aortic regurgitation and left ventricular remodeling after transcatheter aortic valve implantation: A serial cardiac magnetic resonance imaging study. Circulation. Cardiovascular interventions. 2013, 6: 476-483. 10.1161/CIRCINTERVENTIONS.112.000115.

    Article  PubMed  Google Scholar 

  24. Mohty D, Dumesnil JG, Echahidi N, Mathieu P, Dagenais F, Voisine P, Pibarot P: Impact of prosthesis-patient mismatch on long-term survival after aortic valve replacement: Influence of age, obesity, and left ventricular dysfunction. Journal of the American College of Cardiology. 2009, 53: 39-47. 10.1016/j.jacc.2008.09.022.

    Article  PubMed  Google Scholar 

  25. Walther T, Rastan A, Falk V, Lehmann S, Garbade J, Funkat AK, Mohr FW, Gummert JF: Patient prosthesis mismatch affects short- and long-term outcomes after aortic valve replacement. European journal of cardio-thoracic surgery: official journal of the European Association for Cardio-thoracic Surgery. 2006, 30: 15-19. 10.1016/j.ejcts.2006.04.007.

    Article  Google Scholar 

  26. Howell NJ, Keogh BE, Barnet V, Bonser RS, Graham TR, Rooney SJ, Wilson IC, Pagano D: Patient-prosthesis mismatch does not affect survival following aortic valve replacement. European journal of cardio-thoracic surgery: official journal of the European Association for Cardio-thoracic Surgery. 2006, 30: 10-14. 10.1016/j.ejcts.2006.03.046.

    Article  Google Scholar 

  27. Mascherbauer J, Rosenhek R, Fuchs C, Pernicka E, Klaar U, Scholten C, Heger M, Wollenek G, Maurer G, Baumgartner H: Moderate patient-prosthesis mismatch after valve replacement for severe aortic stenosis has no impact on short-term and long-term mortality. Heart. 2008, 94: 1639-1645. 10.1136/hrt.2008.142596.

    Article  CAS  PubMed  Google Scholar 

  28. Fuster RG, Montero Argudo JA, Albarova OG, Sos FH, Lopez SC, Codoner MB, Buendia Minano JA, Albarran IR: Patient-prosthesis mismatch in aortic valve replacement: Really tolerable?. European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery. 2005, 27: 441-449. 10.1016/j.ejcts.2004.11.022. discussion 449

    Article  Google Scholar 

  29. Tasca G, Brunelli F, Cirillo M, DallaTomba M, Mhagna Z, Troise G, Quaini E: Impact of valve prosthesis-patient mismatch on left ventricular mass regression following aortic valve replacement. The Annals of thoracic surgery. 2005, 79: 505-510. 10.1016/j.athoracsur.2004.04.042.

    Article  PubMed  Google Scholar 

  30. Tasca G, Brunelli F, Cirillo M, Amaducci A, Mhagna Z, Troise G, Quaini E: Mass regression in aortic stenosis after valve replacement with small size pericardial bioprosthesis. The Annals of thoracic surgery. 2003, 76: 1107-1113. 10.1016/S0003-4975(03)00823-3.

    Article  PubMed  Google Scholar 

  31. Freed DH, Tam JW, Moon MC, Harding GE, Ahmad E, Pascoe EA: Nineteen-millimeter prosthetic aortic valves allow normalization of left ventricular mass in elderly women. The Annals of thoracic surgery. 2002, 74: 2022-2025. 10.1016/S0003-4975(02)04079-1.

    Article  PubMed  Google Scholar 

  32. Tasca G, Brunelli F, Cirillo M, Dalla Tomba M, Mhagna Z, Troise G, Quaini E: Impact of the improvement of valve area achieved with aortic valve replacement on the regression of left ventricular hypertrophy in patients with pure aortic stenosis. Ann Thorac Surg. 2005, 79: 1291-1296. 10.1016/j.athoracsur.2004.09.002. discussion 1296

    Article  PubMed  Google Scholar 

  33. Pibarot P, Dumesnil JG: Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention. Journal of the American College of Cardiology. 2000, 36: 1131-1141. 10.1016/S0735-1097(00)00859-7.

    Article  CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Cristina Gavina.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ contribution

CG, AG, CA, RH, ALM, FRG take responsibility for all aspects of the reliability and freedom from bias of the data presented and their discussed interpretation. All authors read and approved the final manuscript.

Electronic supplementary material

12947_2014_536_MOESM1_ESM.doc

Additional file 1: Table S1: Correlation of the presence of patient prosthesis mismatch (EAOI ≤ 0.85 cm2) and changes in indexed 3D volumes and left ventricular mass. (DOC 35 KB)

12947_2014_536_MOESM2_ESM.doc

Additional file 2: Table S2: Correlation of the presence of aortic regurgitation at 6 months and changes in indexed 3D volumes and left ventricular mass. (DOC 36 KB)

Authors’ original submitted files for images

Below are the links to the authors’ original submitted files for images.

Authors’ original file for figure 1

Authors’ original file for figure 2

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Gavina, C., Gonçalves, A., Almeria, C. et al. Determinants of clinical improvement after surgical replacement or transcatheter aortic valve implantation for isolated aortic stenosis. Cardiovasc Ultrasound 12, 41 (2014). https://doi.org/10.1186/1476-7120-12-41

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1476-7120-12-41

Keywords